X-ray energy dependence: Simulates relative light output of Photostimulatable Phosphor Plate (PSP) within +/- 3% over kVp range of 60 to 120 kVp and a patient equivalent thickness range of 5 to 35 cm (within specified operating rates).
Digital range: Computed Radiography Light Units; CRLU (AEC#); 0 to 655.35
Maximum CRLU Rate: 2500/sec (approx 250 mR/sec entrance exposure rate)
Power On/Off: Manual switch
Controls: Manually using the Power/Reset button.
Functions: Measures CRLU (AEC#); Converts CRLU to CR manufacturers specific CR plate reader light exposure index value (EI); user selectable. Calculates cassette input exposure values for various X-ray beam conditions (exposure in mR plus backscatter)
Power requirements: 2 x 9Volt Batteries in Interface box.
Typical battery life: Greater than 8 hrs.
Operating environment: 59 degrees to 95 degrees F (15 to 35 degrees C)
X-ray beam filter: 1.5 mm copper (B152-110); 6 inches by 6 inches complete with velcro straps to attach to X-ray tube collimator housing
Electronic Cassette: Dimensions: 10 x 12 x 0.5 inches (24 x 30 x 1.3 cm) Weight: 3.9 lb (1.8 kg)
Software: Microsoft Excel Template and Instruction Manual on included CD.